Erbe USA Inc recalls ERBECRYO 2 Cryosurgical Unit

The stated reason: A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the re

Illustration: a small clinic exam room with an exam table, a stethoscope and a clipboard
Illustration · The Lantern

Erbe USA Inc began a medical device recall on June 26, 2026 whose reported distribution included Ohio. FDA

What the record shows

FDA graded it Class II — use may cause temporary or medically reversible health consequences. Quantity affected: 1126 units. The recall is listed as ongoing. FDA

1 source

Context

This is one of 21 records of this type (Class Ii) from this source in the year preceding it. FDA

The Lantern holds 60 records from this source in total, and 2882 for New Carlisle across every source. FDA

1 source

By the numbers

  • Date of record: June 26, 2026 FDA
  • Recall number: Z-2938-2026 FDA
  • Recalls in this action: 1 FDA
  • Recalling firm: Erbe USA Inc FDA
  • Product: ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000 FDA
  • Reason for recall: A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he… FDA
1 source
How we know this:

Assembled mechanically from the source filing and The Lantern's own record database — no AI-written prose. Every figure above appears verbatim in the cited record, which is linked on this page. FDA

Sources

1Erbe USA Inc recalls ERBECRYO 2 Cryosurgical UnitFDA Recalls — Distributed in Ohio · Jun 26, 2026 · the filing on this pagefda.gov (opens the agency's page)

The source filing

Recall numberZ-2938-2026
Recalls in this action1
Recalling firmErbe USA Inc
ProductERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
Reason for recallA strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
ClassificationClass II — use may cause temporary or medically reversible health consequences
Recall typemedical device
Quantity1126 units
StatusOngoing
Voluntary or mandatedVoluntary: Firm initiated
Initiated20260626
Reported20260819
DistributionUS Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexi
Firm locationMarietta, GA, United States