← The Lantern · New Carlisle
health · class ii

Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7

Model Number: MEXL-3020/B7. beginning October 16, 2025, in a medical device recall whose distribution included Ohio.

Illustration: a pharmacy counter with pill bottles, a blister pack and one bottle set aside
Illustration · The Lantern

Canon Medical System, USA, INC. began a medical device recall on October 16, 2025 whose reported distribution included Ohio. FDA

What the record shows

recalled VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. beginning October 16, 2025, in a medical device recall whose distribution included Ohio. The stated reason: Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment. FDA graded it Class II — use may cause temporary or medically reversible health consequences. Quantity affected: 2 units. The recall is listed as ongoing. FDA

1 source

Context

This is one of 21 records of this type (Class Ii) from this source in the year preceding it. FDA

The Lantern holds 60 records from this source in total, and 2879 for New Carlisle across every source. FDA

1 source

By the numbers

  • Date of record: October 16, 2025 FDA
  • Recall number: Z-0327-2026 FDA
  • Recalls in this action: 1 FDA
  • Recalling firm: Canon Medical System, USA, INC. FDA
  • Product: VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. FDA
  • Reason for recall: Affected system contains components that do not meet the required distance between electrodes for compliance with the… FDA
1 source
How we know this:

Assembled mechanically from the source filing and The Lantern's own record database — no AI-written prose. Every figure above appears verbatim in the cited record, which is linked on this page. FDA

Sources

1Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7FDA Recalls — Distributed in Ohio · Oct 16, 2025 · the filing on this pagefda.gov (opens the agency's page)

The source filing

Recall numberZ-0327-2026
Recalls in this action1
Recalling firmCanon Medical System, USA, INC.
ProductVANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Reason for recallAffected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
ClassificationClass II — use may cause temporary or medically reversible health consequences
Recall typemedical device
Quantity2 units
StatusOngoing
Voluntary or mandatedVoluntary: Firm initiated
Initiated20251016
Reported20251105
DistributionUS distribution to Ohio and Nevada
Firm locationTustin, CA, United States
View the original at fda.gov